Syringe Pump vs HPLC Pump: Which Suits Low Flow Dosing? Pulsation, Pressure, Volume?

Which pump gives a small, steady flow? And which one keeps going when the line pushes back? This syringe pump vs HPLC pump guide uses the maker's own specs, so you can check every number. Where the maker gives no data, it uses simple pump basics.

A syringe pump is a good fit for slow dosing into open paths that do not push back much. An HPLC pump is a good fit for high back pressure and runs that never stop. Take Infusetek's CFSP-I/CU-I. It runs from 0.06 μL/min to 70 mL/min. Its liquid path limit is 0.68 MPa.

Below, you will find how each pump works and how they compare on pulsation, pressure and volume. A job table, a test plan and six short answers come last.

What Does Low Flow Dosing Need From a Pump?

Low flow dosing means giving a small amount of liquid at a slow, steady speed. Microliter flow means a very small flow, counted in millionths of a liter per minute. A good pump must handle three things: pulsation, pressure and volume.

What Is the Purpose of a Syringe Pump?

A syringe pump moves a small amount of liquid at a set speed. DK INFUSETEK lists uses in microfluidics, flow chemistry, battery work, coating, electrospinning and dispensing. Its pumps are for labs and factories. They are not for medical use.

Syringe Pump vs HPLC Pump at a Glance

A syringe pump is suitable for gradual dosing into low-resistance channels. An HPLC pump is good for high back pressure and runs that never stop. The table compares them side-by-side.

Syringe Pump and HPLC Pump Compared

Factor

Syringe pump (maker data)

HPLC pump (general, no numbers)

Flow and volume

ISPLab01-Pro: 0.008 nL/min to 165.871 mL/min, set by syringe size

Fed from a solvent bottle, so run time follows the bottle

Pulsation

Plunger drive. CFCU-I is a low-pulsation design

A reciprocating piston and check valves cause ripple. A pulse dampener or dual-piston design smooths it

Pressure

0.68 MPa max liquid path (CFSP-I/CU-I, SP60-8A)

Built for high back pressure from packed columns

Run time

Standard pumps stop when the syringe is empty. Continuous flow models do not

Runs as long as solvent is supplied

For slow flow into easy paths, pick a syringe pump. That is the answer in the syringe pump vs HPLC pump choice. When the path pushes back, pick an HPLC pump.

How Does a Syringe Pump Make Slow, Steady Flow?

A syringe pump pushes a plunger at a constant pace. A lead screw driven by a stepper motor. The screw pushes the plunger down the syringe.

The flow rate equals the speed of the plunger times the inner area of the syringe. The ISPLab01-Pro has a linear speed of 1 μm/min to 132 mm/min. Its stroke resolution is 0.039 µm/µstep.

The narrower the syringe hole, the lower the flow. This is seen in the row of 0.5 μl micro syringes below.

How to Calculate Flow Rate in a Syringe Pump?

Multiply the interior area of the syringe by plunger rate. Take a 1 mL plastic syringe with a 4.70 mm bore. Its area is about 17.35 mm². At 10 mm/min, the flow is about 173 μL/min. Remember, 1 mm³ is the same as 1 μL.

Syringe Size and Flow Range (ISPLab01-Pro)

Syringe

Inner diameter

Reference flow range

Micro 0.5 μL

0.10 mm

0.008 nL/min to 1.0296 μL/min

Plastic 1 mL

4.70 mm

0.0173 μL/min to 2.2889 mL/min

Plastic 150 mL

40.00 mm

1.2566 μL/min to 165.8712 mL/min

How Does an HPLC Pump Move Liquid?

A reciprocating piston in an HPLC pump propels the liquid. HPLC stands for high-performance liquid chromatography. The piston draws in solvent and pushes it out.

Inlet and exit check valves cause the liquid to go one direction. Every stroke is a little push. So the flow goes up and down a little bit. This ripple is known as a pulse.

Each thrust has something absorbed by a pulse dampener. A dual piston design allows one piston to refill while the other is pushing. This section describes how these pumps function in general. It's not manufacturer data.

Pulsation: Which Pump Gives Smoother Low Flow?

In the syringe pump vs HPLC pump pulsation test, neither pump is free of pulses by default. Smooth flow depends on the design and the setup. A plunger drive does not ripple stroke by stroke while it pushes. But a real setup can still pulse.

Why Do Piston Pumps Ripple?

In strokes, liquid is propelled by a piston that reciprocates. Every stroke opens and closes the check valves. Each stroke is little when the flow is low. Therefore, the flow is greatly affected by even the smallest wave. Subdue it using a pulse dampener.

Where Can Syringe Pumps Still Pulse?

When the syringe is filled or the valve is turned on, the pump might pulse. Pulsing or delay may also be caused by air in the syringe or tubing. It may be fixed by priming the line again.

What Helps Low Flow Stability?

CFCU-I: two actuators work together, with a compensation and calibration algorithm. It is made for low-pulsation, steady flow.

CFSP-I/CU-I: closed-loop control and overload protection.

SPM: a back distance setting (0.01 to 5 mm) that clears bubbles out of the syringe.


The site gives no pulsation number, so this article shows none.

Pressure: Can a Syringe Pump Push Against a Packed Column?

The top liquid path pressure for CFSP-I/CU-I and SP60-8A is 0.68 MPa (about 6.8 bar). For more pressure on the CFSP-I/CU-I, contact the maker. The CFCU-I also lists 0.68 MPa. It has optional upgrades for higher pressure.

What Sets the Pressure Rating?

Borosilicate glass syringe barrel

Stainless steel plunger rod

PTFE plunger seal

PCTFE and PTFE valve (the CFCU-I also lists ceramic and PEEK valves)


The ISPLab01-Pro pushes with a force of 12 to 45 kgf. That is force, not pressure, so this article does not convert it. For paths that push back hard, an HPLC pump is the better fit.

Volume: How Long Can Each Pump Run?

A syringe pump runs until the syringe is empty. An HPLC pump runs as long as solvent keeps coming. So the syringe size sets the batch size.

The CFSP-I/CU-I takes syringes from 50 μL to 25 mL. The ISPLab01-Pro takes 0.5 μL to 150 mL. Continuous flow models do not have to stop. These are the CFSP-I/CU-I and the CFCU-I. Their dispensing volume range is 0.1 μL to 25 mL.

Syringe Size vs Flow Limits (CFSP-I/CU-I)

Syringe

Minimum flow

Maximum flow

50 μL

0.0625 μL/min

608.37 μL/min

1.0 mL

1.25 μL/min

12.16 mL/min

10 mL

12.5 μL/min

70.00 mL/min

25 mL

31.25 μL/min

55.00 mL/min

The maker tested these limits with pure water. The inlet tube was under 1 m long, with a 1.6 mm inner diameter.

How Do You Read Accuracy Numbers?

Each accuracy number belongs to one model and one test rule. Mixing them can make low flow stability look better than it is.

Accuracy by Model and Test Rule

ISPLab01-Pro: error of at most ±0.3% when the stroke is at least 30% of the maximum.

SPM: error of at most ±0.5% under the same 30% rule.

CFSP-I/CU-I: stroke control error of at most ±0.1% at 30% stroke or more. This is stroke control, not flow accuracy.

SP60-8A: repeatability <0.2% and accuracy <1.0%. This is at full stroke with deionized water.


Which Pump Fits Which Low Flow Job?

A syringe pump fits most microliter flow jobs. High back pressure is the one case where an HPLC pump fits better.

Low Flow Jobs and Pumps

Job

Pump

Why

Microfluidics and electrospinning

ISPLab01-Pro

0.008 nL/min to 165.871 mL/min; the site has an electrospinning case note

LNP preparation

SPM/4*ZU-I

1 to 8 independent units; the site has an LNP case note

Flow chemistry and dosing without a stop

CFSP-I/CU-I

Continuous flow, 0.06 μL/min to 70 mL/min

Multi-channel pipetting and dilution

SP60-8A

8 channels with a rotary valve

Packed column separation

An HPLC pump

Built for high back pressure

Is a Syringe Pump the Same as an Infusion Pump?

No, they are not the same. A syringe pump pushes a plunger. It suits small volumes and very slow flow. An infusion pump moves liquid from a bag through tubing. It suits large volumes and long delivery. A syringe holds a small, fixed amount. A bag holds much more. Each one fits a different job. Pick the pump that fits your volume and run time. DK INFUSETEK's pumps are for labs and factories. This article keeps all uses non-medical, as the site does.

How Do You Run a Low Flow Test Before the Real Sample?

A water test shows if your setup holds low flow stability. Run it before the real sample goes in.

  1. Pick the syringe, or type in its inner diameter.

  2. Fill the syringe and push out the air.

  3. Load the syringe and clamp it.

  4. Prime the line until a steady drip shows.

  5. Run a water test with at least 30% of the stroke.

The ISPLab01-Pro max stroke is 140 mm. So its test moves at least 42 mm. This applies to the ISPLab01-Pro only.

What Affects the Cost of a Low Flow Pump?

This article gives no prices because the site lists none. The site names these cost drivers: flow range and precision, number of channels, valve material and control options. For price and delivery time, email sales@infusetek.com. The site gives no delivery time, so none is given here.

About DK INFUSETEK (Duko Industrial Technology (Shanghai) Co., Ltd.)

DK INFUSETEK has its headquarters and R&D center in Shanghai. Its production base is in northern China. The company designs, customizes and makes devices for micro-fluid handling.

It has five product lines: laboratory, continuous flow, modular, industrial and accessories. Support includes datasheet PDFs, video tutorials and online consultation. It also posted an ACHEMA 2024 notice for June 10 to 14 in Frankfurt.

FAQs

Can an HPLC pump deliver very low flow rates accurately?

Yes, often, providing the flow is within the maker's range. It becomes tougher on the bottom end. Each stroke is modest there thus the ripple counts more. First try the pump with your own fluids. For gradual dosing to simple pathways, check out the syringe pump alternatives at infusetek.com.

How do you calculate flow rate in a syringe pump?

Multiply the plunger speed by the interior area of the syringe . Use a 1 mL plastic syringe with 4.70 mm bore. The flow at 10 mm/min is roughly 173 μL/min. The total length of antenna is 21.1 mm. A microliter is a cubic millimeter.

What is the purpose of a syringe pump?

A syringe pump delivers tiny volumes of fluid at a programd rate. It is used in labs and industries for microfluidics, flow chemistry, battery work, coating, electrospinning and dispensing. DK INFUSETEK states their pumps are for these reasons, not medical usage.

What is the difference between a syringe pump and an infusion pump?

A plunger is pushed by a syringe pump. Suitable for low quantities and extremely low flow. An infusion pump pushes fluids from a bag via tubes. Good for big lots and extended delivery. Pick the one that matches your volume and run time.

Can a syringe pump run without stopping?

Yes, but only a continuous flow model can. The CFSP-I/CU-I covers 0.06 μL/min to 70 mL/min without a stop. The CFCU-I is a continuous flow model too. A standard syringe pump must stop when its syringe is empty. Check which model fits your flow range first.

What pressure can a continuous flow syringe pump handle?

The CFSP-I/CU-I can handle up to 0.68 MPa in the liquid path. That is about 6.8 bar. For a higher pressure rating, contact the maker. The CFCU-I also lists 0.68 MPa. It has optional upgrades for more. Ask the maker which option fits your job.

Conclusion

Pick a syringe pump for microliter flow into easy paths. Pick an HPLC pump when the job has high back pressure. When you weigh a syringe pump vs HPLC pump, look at the pressure rating first. Then look at how much liquid you need for each run. 

Duko Industrial Technology (Shanghai) Co., Ltd. shares specs, syringe tables and accuracy rules for each model. Read them before you buy. Match each number to its own model. To start, get a datasheet, chat online, or email sales@infusetek.com.

 
logo
About Us

Company Profile
Product

Laboratory Syringe Pump
Continuous flow syringe pump
Modular Syringe Pump
Industrial Syringe Pump
Accessories
Application

Life Science
Microfluidic
University Laboratory
Electrospinning
Support

Download

DK INFUSETEK

Add:No.355, Building B, Hongqiao World Center, 
Yinggang Dong Road, Qingpu District, 
Shanghai 201700 , CHINA
Tel: 0086-312-5958380 / 0086-19903128869
Web: www.infusetek.com
Email: sales@infusetek.com
    Contact
    度科工业科技(上海)有限公司版权所有 沪ICP备2020030748号-1

    We use cookies

    Our website uses cookies to enhance user experience, personalise content and show you more relevant adverts about our great products, also to analyze our website traffic.